Warning
Description
Temperature is not
The temperature has
calibrated.
not been calibrated.
Calibration is in
A calibration was
progress.
started but not
completed.
9.3 Error list
When an error occurs, the reading on the measurement screen flashes and all outputs are held when
specified in the Controller > Outputs menu. The screen changes to red. The diagnostics bar shows
the error. Push on the diagnostic bar to show the errors and warnings. As an alternative, push the
main menu icon, then select Notifications > Errors.
A list of possible errors is shown in
Error
ADC failure
Temperature sensor is
missing!
Section 10 Replacement parts and accessories
For SC4500 Controller parts and accessories, refer to the replacement parts and accessories section
of the SC4500 Controller User Manual on the manufacturer's website.
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Sample conditioning solution
Description
Diisopropylamine (DIPA)
Ammonia solution, 28-30%
(NH
OH)
4
Ammonia solution, 28-30%
(NH
OH)
4
6
Not available in all regions.
24 English
Table 3 Warning list (continued)
Resolution
Calibrate the temperature.
Return to the calibration.
Table
4.
Table 4 Error list
Description
The analog to digital
conversion failed.
The temperature sensor is
missing.
W A R N I N G
6
99%
Resolution
Make sure that the ultrapure amperometry module is fully
inserted into the module connector in the controller.
Power off and power on the controller. Contact technical
support.
Examine the cable connection at the working electrode.
Clean the working electrode with a soft cloth. Contact
technical support.
Quantity
1 L
500 mL
2.5 L
Item no.
2834453
10649
10604