Temperature Measurement (Optional) - Gima OXY 110 Manuel De L'utilisateur

Masquer les pouces Voir aussi pour OXY 110:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

 Remove nail polish or other cosmetic products from the fingernail.
 The fingernail should be of normal length.
 The SpO
sensor cannot be immersed into water, liquid or cleanser.
2
 The SpO
sensor can be repeatedly used. Please clean and disinfect before reuse.
2
 Connector with the label "SpO
connector with the label "TEMP" can only be connected with the temperature
probe.

3.2 Temperature Measurement (optional)

The infrared temperature probe is a delicate transducer. To operate please follow these
steps and procedures. Failure to accurately operate may cause damage to the probe.
The infrared temperature probe is as shown in figure 3.2.
Please place the infrared temperature probe in a stable ambient temperature for 30
minutes before taking a measurement.
Measuring tip of
temperature probe
Temperature
Probe Cover
Small screen
Power on/off key
Battery cover
Figure 3.2 the infrared temperature probe
Operation procedure:
1. Connect the infrared temperature probe to the connector on the upper side of
device marked with "TEMP".
2. When the screen shows as the big display screen in figure 3.2 and the temperature
unit "°C" is blinking, the user can begin to take the measurement.
3. Insert the tip of the temperature probe into the earhole and press the measuring key
to start the measurement. A short beep means the measurement has finished and
the result will be displayed on the big display screen on temperature probe and the
display screen of the Oximeter.
ENGLISH
" can only be connected with SpO
2
probe, and
2
Measuring key
Big display screen
46

Publicité

Table des Matières
loading

Table des Matières