The earth terminal of the casing must be connected to a reliable grounding (see the universal ground
symbol on the housing).
Connect the power cord to a grounded socket.
The air flow speed of the worktable should be adjusted to the best state while dispatching from the
factory. It should be adjusted to the low-grade location in the initial stage, and generally improve in half
a year.
Device use
If the worktable is newly installed or has not been used for a long time, clean it and the surrounding
environment with dust catcher or the tool that will not produce fiber before using, and then use
medicine sterilization or the ultraviolet sterilisation method to deal with sterilization.
Start the appliance 5 minutes in advance and turn on the UV-light.
Shut off the sterilisation light after 30 minutes, and then turn on the air blower.
The operation can be started 20 minutes later, turn on the fluorescent lamp while operating.
Maintenance
a)
Unplug the mains plug and allow the device to cool completely before each cleaning, adjustment or
replacement of accessories, or if the device is not being used.
Wait for the rotating elements to stop.
b)
Store the unit in a dry, cool place, free from moisture and direct exposure to sunlight.
c)
Do not spray the device with a water jet or submerge it in water.
d)
Do not allow water to get inside the device through vents in the housing of the device.
e)
Clean the vents with a brush and compressed air.
f)
Use a soft cloth for cleaning.
g)
Do not use sharp and/or metal objects for cleaning (e.g. a wire brush or a metal spatula) because they
may damage the surface material of the appliance.
h)
HEPA-filter lifetime:
When the fan had been running less efficiently and not reaching the air flow speed it is indicating that
the user needs to change the filter. While changing the high-efficient filter, open the top cover, pay
attention to the arrow sign on the filter, the direction of arrow indicates the laminar flow. Check
carefully while changing the filter, make sure the frame is sealed well all around; otherwise, the filter
result will be influenced.
i)
UV-sterilising lamp lifetime:
When the UV-sterilising lamp had been on for about 1000 hours it needs to be replaced.
j)
Sterilise the environment periodically (usual once a week), Use the gauze with alcohol for sterilisation.
At the same time, keep the surface clean. Otherwise it will influence sterilisation result.
Troubleshooting
Issue
No power supply.
The air fan does not work.
The light does not work
Cannot reach required air
flow speed
Disposing of used devices
Do not dispose of this device in municipal waste systems. Hand it over to an electric and electrical device recycling
and collection point. Check the symbol on the product, instruction manual and packaging. The plastics used to
EN
Possible cause
1. The plug and socket are poorly
connected.
2. The fuse is burnt.
The control does not work or the air
fan is broken down.
The light socket is broken down.
The light tube is burnt.
The high-efficient filter wears out.
Solution
1. Adjust the plug and
socket, or check the circuit.
2. Change the fuse.
Change the control panel or
check the air fan.
Adjust the lamp socket or
change the light tube.
Change the filter.