Rated
Separation distance according to
maximum
frequency of transmitter (m)
output
150 kHz
80 MHz to
power of
to 80 MHz
800 MHz
transmitter
d=1.2
P
d=1.2
(W)
0.01
0.12
0.12
0.1
0.38
0.38
1
1.2
1.2
10
3.8
3.8
100
12
12
For transmitters rated at maximum output power not
listed above, the recommended separation distances (d)
in meters (m) can be estimated using the equation ap-
plicable to the frequency of the transmitter, where (P) is
the maximum output power rating of the transmitter in
watts (W) according to transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation dis-
tance for the higher frequencyrange applies.
NOTE 2
These guidelines may not apply in all situ-
ations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and
people.
800 MHz to
2.5 GHz
P
d=2.3
P
0.23
0.73
2.3
7.3
23
44