Clean the cuvette holder after each day of use.
1. Turn off the analyzer. Place the cuvette mov-
ing arm in loading position.
2. Remove cuvette holder by lifting it straight
up.
3. Clean the cuvette holder with alcohol
(20-70%) or mild detergent. NOTE: Do not
autoclave. If the optical parts are stained an
error code is displayed. Clean according to
the Maintenance/Optical Parts section.
4. Wait 15 minutes before replacing the cuvette
holder.
5. Place the cuvette moving arm in loading
position before replacing the cuvette holder.
Clean the cover with alcohol (20-70%) or a
mild detergent.
Only the WBC cuvette holder is to be used
!
with the WBC analyzer. The WBC cuvette holder
is marked WBC.
CFR
Nettoyer le support de cuvette après chaque
utilisation.
1.
Éteindre l'analyseur. Mettre le bras mobile
de la cuvette en position de charge.
2.
Retirer le support de cuvette en le soulevant
sans l'incliner.
3.
Nettoyer le support de cuvette avec de
l'alcool (20 à 70 %) ou un détergent doux.
REMARQUE : Ne pas stériliser. Si des pièces
optiques sont tachées, un code d'erreur
s'affiche. Nettoyer conformément à la
section Entretien/Pièces optiques.
4.
Attendre 15 minutes avant de remettre
le support de cuvette en place.
5.
Placer le bras mobile de la cuvette en position
de charge avant de remplacer le support
de la cuvette. Nettoyer le couvercle avec de
l'alcool (20 à 70 %) ou un détergent doux.
!
Seul le support de cuvette WBC doit être
utilisé avec le WBC Analyzer. WBC est marqué
sur le support de cuvette WBC.
27