Symptom
Err60
Err61
Err62
WbC
LLL
HHH
Explanation
General hardware error.
Self test error during start up of analyzer.
1. Blanking test failed.
2. Optical parts dirty
Empty cuvette holder.
1. Measured value is below 0.3 x10
(300/mm
, 300/µL).
3
2. Empty microcuvette, not filled with
sample.
Measured value exceeds 30.0 x10
(30000/mm
3
, 30000/µL).
Action
a. Turn off the analyzer and turn it on again after
30 seconds. Take a new microcuvette and repeat the
measurement, as described in the Measuring section.
b. If the problem continues, the analyzer needs service.
Contact your local distributor.
a. Turn off the analyzer and turn it on again after
30 seconds. Take a microcuvette and perform the
measurement, as described in the Measuring section.
b. If the problem continues, the analyzer needs service.
Contact your local distributor.
1. Turn off the analyzer and turn it on again after
30 seconds. Take a new microcuvette and repeat the
measurement, as described in the Measuring section.
2a. Turn off the analyzer and clean the optical parts as
described in the Maintenance section.
2b. If the problem continues, the analyzer needs
service. Contact your local distributor
Take a microcuvette and perform the measurement, as
described in the Measuring section.
9
/L
1. NOTE! Results exceeding the LLL limit should be
verified with a suitable laboratory method and be
questioned as to the pathological condition of the
patient.
2. Take a new microcuvette and repeat the
measurement, as described in the Measuring section
and make sure that the microcuvette is filled with
sample.
/L
NOTE! Results exceeding the HHH limit should be
9
verified with a suitable laboratory method and be
questioned as to the pathological condition of the
patient.