Interference Studies
Interference studies were performed according to CLSI
Document EP7-A2. The following substances have not
been found to interfere with the HemoCue WBC system.
The highest concentration or percentages tested is
referred to in brackets.
Acetaminophen (20 mg/dL), ascorbic acid (6 mg/dL),
ibuprofen (50 mg/dL), caffeine (308 µmol/L), creatinine/
urea (30 mg/dL/500 mg/dL), salicylic acid (60 mg/dL),
tetracycline (30 mg/dL), unconjugated bilirubin
(342 µmol/L), conjugated bilirubin (380 µmol/L), dex-
tran 40 (30 g/L), osmolality (167 mmol/L), HbCO (30 %),
methemoglobin (50 %), thrombocytes (2000 x10
lipemia (intralipid 3100 mg/L) which corresponds to
triglycerides (approximately 10.5 mmol/L), hemoglobin
(30–250 g/L), reticylocytes (258 x10
leukocytes and platelet clumping have not been found
to interfere. pH values between 6.8–8.0 do not interfere
with the system.
Technical Specifications
Dimensions: 179 x 133 x 121 mm (7.05 x 5.24 x 4.76 inches)
Weight: 600 g (1.32 pounds) (with 6 AA batteries in-
stalled)
Power adapter: CE marked
Only use adapters, as listed under Adapters.
Pollution degree: 2
Overvoltage category: II
Atmospheric pressure: 700 hPa to 1060 hPa.
Equipment not suitable for use in the presence of flam-
mable mixtures.
The HemoCue WBC system has been tested for electrical
safety and EMC according to the following standards:
• IEC/EN 61010-2-101, 2002 (First Edition), IEC 61010-1,
2001 (Second Edition) and EN 61010-1, CAN/CSA-C22.2
no.61010-1, 2004 (Second Edition), UL 61010-1, 2004
(Second Edition), IEC 60601-1-2:2007 (Third Edition)
and EN 60601-1-2:2007, IEC 61326-1:2005 and EN 61326-
1:2006, and IEC 61326-2-6:2005 (First Edition) and EN
61326-2-6:2006
Essential Performance
The HemoCue WBC system is an In-Vitro diagnostic
system designed for quantitative determination of white
g/L),
blood cells (WBC) in capillary or venous whole blood.
9
Recommended separation distance between Portable
L), HbAS, immature
and mobile RF communications equipment and
9
HemoCue WBC Analyzer
The HemoCue systems are intended for use in an electro-
magnetic environment in which radiated RF disturbances
are controlled. The customer or the user of HemoCue
systems can help prevent electromagnetic interference
by maintaining a minimum distance between portable
and RF communications equipment (transmitters) and
HemoCue systems as recommended below, according
to the maximum output power of the communications
equipment.