3. Specifications / 4. Design and function
Process connection
In accordance with the general technical regulations for pressure
gauges (e.g. EN 837-2 "Selection and installation recommendations for
pressure gauges"), plastic, G 1/8 female threads
Operating temperature
Ambient: -30 ... +80 °C
Medium: -16 ... +50 °C
Temperature effect
When the temperature of the measuring system deviates from the
reference temperature (+20 °C): max. ±0.5 %/10 K of full scale value
Ingress protection
IP 54 per EN 60529 / lEC 529, optionally IP 65
EMC Directive
Per EN 61000-6-4 and EN 61000-6-2 (cable length max 30 m)
For further specifications see WIKA Data Sheet PM 07.40, PV 17.40
and the order documentation.
4. Design and function
Description
Nominal size 110 mm
■
The instruments measure the pressure by means of diaphragms
■
(silicone)
The measuring characteristics are in accordance with the EN 837-3
■
standard and ventilation and air conditioning (VAC) regulations
Scope of delivery
Cross-check scope of delivery with delivery note.
WIKA operating instructions air2guide models A2G-10, A2G-15
GB
7