Guidance and Manufacturer's Declaration - Electromagnetic Immunity:
The Concentrator is intended for use in the electromagnetic environment specified below. The user of the Concentrator
should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz
to 80 MHz
Radiated RF
IEC 61000-4-3
3V/m
80 MHz
to 2.5 GHz
Electrostatic
± 6 kV contact
discharge (ESD)
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power
transient/burst
supply lines
± 1 kV for input/
EC 61000-4-4
output lines
Surge
± 1 kV line(s) to
line(s)
IEC 61000-4-5
± 2 kV line(s) to
earth
<5% U
Voltage dips, short
(>95% dip
T
in U
interruptions and
) for 0.5 cycle
T
voltage variations
40% U
on power supply
(60% dip in
T
U
input lines
) for
T
5 cycles
IEC 61000-4-11
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip
T
in U
) for 5 sec
T
Power frequency
3 A/m
(50/60 Hz)
magnetic
field
IEC 61000-4-8
Compliance
Electromagnetic
Level
Environment - Guidance
Portable and mobile RF communications equipment should be used no closer
to any part of the device, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d=1.2√
150 kHz to 80 MHz
P
d=1.2√
80 MHz to 800 MHz
P
d=2.3√
800 MHz to 2.5 GHz
P
3 Vrms
Where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
3V/m
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
frequency range
Interference may occur in the vicinity of equipment marked with the following
symbol:
± 6 kV contact
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
± 8 kV air
± 2 kV for power
Mains power quality should be that of a typical commercial or hospital
supply lines
environment.
± 1 kV for input/
output lines
± 1 kV line(s) to
Mains power quality should be that of a typical commercial or hospital
line(s)
environment.
± 2 kV line(s) to
earth
<5% U
Mains power quality should be that of a typical commercial or hospital
(>95% dip
T
in U
environment. If the user of the [ME EQUIPMENT or ME SYSTEM] requires
) for 0.5 cycle
T
continued operation during power mains interruptions, it is recommended
40% U
that the [ME EQUIPMENT or ME SYSTEM] be powered from
(60% dip in
T
U
an uninterrupted power supply or a battery.
) for
T
5 cycles
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip
T
in U
) for 5 sec
T
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical hospital or home environment.
, should be less than the compliance level in each
a
.
b
41