Télécharger Imprimer la page

Inogen ING-BA-401 Manuel De L'utilisateur page 30

Publicité

Low Priority Alerts (continued)
Message Display & Text
Check Battery
Oxygen Low
Remove Battery
to Cool
Service Soon
Sensor Fail
26
Condition/Action/Explanation
Battery error has occurred. Check the connection of your
battery and ensure that it is properly attached and latched
on concentrator. If battery error recurs with same battery,
stop using the battery and switch to a new battery or
remove battery and operate concentrator using external
power supply.
Concentrator is producing oxygen at a slightly low level
(<82%) for a period of 10 minutes. If condition persists,
contact your equipment provider.
Battery has exceeded its charging temperature and charging
has stopped. The battery will not charge while this alert
is present but will begin to charge when the battery
temperature returns to the normal operating range.
If battery charging is desired sooner, remove the battery
from the concentrator and allow it to cool in an open area for
approximately 10-15 minutes. Then, re-insert the battery into
the Inogen One® G4. If the problem still persists, contact your
equipment provider.
The concentrator requires servicing at the earliest
convenience. The concentrator is operating to specification
and may continue to be used. Contact your equipment
provider to arrange for service.
The concentrator's oxygen sensor has malfunctioned.
You may continue to use the concentrator. If the condition
persists, contact your equipment provider.

Publicité

loading

Ce manuel est également adapté pour:

One g4