General Instructions
1. Place the Inogen One® G4 in a well ventilated location.
Air intake and exhaust must have clear access.
Locate the Inogen One® G4 in such a way that any
auditory alerts may be heard. Always operate the
Inogen One® G4 in an upright position (see image
for proper orientation).
WARNING Avoid use of the Inogen One® G4 in presence of pollutants, smoke
or fumes. Do not use the Inogen One® G4 in presence of flammable
!
C
anesthetics, cleaning agents or other chemical vapors.
CAUTION
Do not obstruct air intake or exhaust when operating the device.
Blockage of air circulation or proximity to a heat source may lead to
internal heat buildup and shutdown or damage to the concentrator.
CAUTION
The Inogen One® G4 Concentrator is designed for continuous use. For
optimal sieve bed (columns) life, the product should be used frequently.
2. Ensure particle filter is in place.
CAUTION
Do not operate the Inogen One® G4 without the
particle filter in place. Particles drawn into the
system may damage the equipment.
Operating Instructions
Intake
Exhaust
Operating Instructions Chapter 3
This side
UP
Exhaust
11