NOTE
To purchase additional particle filters contact your equipment provider or Inogen.
Output Filter
The output filter is intended to protect the user from inhalation of small particles in the
product gas flow. The Inogen One® G4 includes an output filter conveniently located
behind the removable cannula nozzle fitting.
The output filter may be replaced by the equipment provider or by the owner using the
Output Filter Replacement Kit (RP-404).
The Inogen One G4 Concentrator must be cleaned and disinfected as per the above
instructions for each new patient. No special maintenance needs to be carried out
by the patient. Your equipment provider performs maintenance operations to assure
continued reliable service from your Inogen One G4. The manufacturer's instructions for
the preventative maintenance of the devices are defined in the service manual. All work
must be performed by trained technicians certified by the manufacturer.
DC Input Cable Fuse Replacement
The Cigarette Lighter DC power plug contains a fuse. If the DC input cable is being used
with a known good power source and the unit is not receiving power, the fuse may need
to be replaced.
To replace the fuse, follow these instructions and refer to the photograph below.
1. Remove the tip by unscrewing the retainer. Use a tool if necessary.
2. Remove the retainer, tip and fuse.
3. The spring should remain inside the Cigarette Lighter Adapter housing. If the spring
is removed, make sure to replace the spring first before inserting the replacement
fuse.
32
It may be necessary to clean the particle filter more often in dusty
environments.