9.4 Error messages
Error
message
- - - - -
Maximum load exceeded
- - ol - -
Err 1
Incorrect data input
Err 2
Incorrect time entry
Err 4
Zeroing range exceeded due
to switching-on balance or
pressing
max)
Err 5
Keyboard error
Err 6
Value outside the A/D changer
range
Err 9
Stability display does not
appear
Err 10
Communication error
Err 15
Gravitation error
Err 17
Taring range exceeded
Err 19
Zero point displaced
Fai l h /
Adjustment error
Fai l l
Err P
Printer error
Ba lo /
Battery very low
Lo ba
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
50
Description
(normally 4%
Possible causes
• Unload weighing system or reduce
preload.
• Follow format "yy:mm:dd"
• Follow format "hh:mm:ss"
• Object on the weighing plate
• Overload when zeroing
• Weighing plate not installed
• Damaged weighing cell
• Damaged electronics
• Check the environmental
conditions.
• No data
• Range 0.9 ~ 1.0
• Reduce load
• Remedy:
Adjust / linearize
• Repeat adjustment.
• Check communication parameters
• Recharge battery
SFB-BA-e-1630