 Press
; weighing system is in tolerance weighing mode.
From here evaluation takes place whether the goods to be
weighed are within the two tolerance limits.
Weighing with tolerance range
 Set item weight, see chap. 7.10.
 Tare when using a weighing container.
 Put on goods to be weighed, tolerance control is started. The signal lights
indicate whether the load is within the two set limits.
Load below specified
tolerance
Red signal light
next to „-" ON illuminated
• The tolerance control is not active when the weight is under 20d.
• To delete limits, enter „00000 PCS".
SFB-BA-e-1630
Load within specified
tolerance
Green signal light
next to „„ illuminated
Load exceeds specified
tolerance
Red signal light
next to „+" ON illuminated