• When puncturing, suturing, or incising the tissue near the
sheath, be careful not to damage the sheath. Do not put a clamp
on the sheath or bind it with a thread.
• When inflating a balloon at, or close to, the sheath tip, be sure
not to inflate it inside the distal end of the sheath (Fig. 6). The
radiopaque marker is located approximately 5 mm proximal
to the tip (Fig. 1) and therefore does not actually mark the true
distal tip of the sheath.
8. The side tube may also be used as a continuous infusion site by
connecting an infusion line to the 3-way stopcock.
9. When inserting, manipulating or withdrawing an interventional /
diagnostic device from the sheath, always be sure to hold the
sheath in place.
Fig. 6
CAUTION
• Do not suture the sheath tubing since this may restrict access /
flow through the sheath.
• Do not use a power injector through the side tube and 3-way
stopcock. Excessive leakage may occur through the CCV valve
with high/rapid flow injections such as an injection of contrast
to provide a comprehensive image of the aortic arch.
10. After the intended procedure is completed, remove all of the
devices.
STORAGE
Do not store at extreme temperatures and humidity. Avoid direct
sunlight.
Correct
Incorrect