Résultats Techniques De La Cem - Silk-n ToothWave H7001 Mode D'emploi

Masquer les pouces Voir aussi pour ToothWave H7001:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

17. Résultats techniques de la CEM
EMC requirements
IEC 60601-1-2 Edition 4.0 (2014)
Environment of intended uses:
Home Healthcare Facility Environment
Summary of Test Results
Test
Emission
Conducted emission
Freq. range:
150 kHz – 30 MHz
Radiated emission
Freq. range:
30 MHz – 1 GHz
Voltage changes,
Voltage fluctuations
and Flicker test
Immunity (IEC 60601-1-2 section 8)
Immunity from
Electrostatic
discharge (ESD)
Immunity from
radiated
electromagnetic
fields
Immunity from
Proximity field from
wireless
communications
equipment
Immunity from
power frequency
magnetic field
Standard
FCC Part 15
Subpart B
EN 61000-3-3
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
47
Class/Severity
level
Class B
on 120 Vac mains
Class B
AC mains
8 kV contact
discharges & 15 kV
air discharges
3.0; 10.0 V/m; 80
MHz 2.7 GHz, 80%
AM,
1 kHz
List of frequencies,
from 9 V/m up to
28 V/m, PM (18 Hz
or 217 Hz), FM 1
kHz
30 A/m @ 50 Hz
Test result
Complies
Complies
Complies
Complies
Complies
Complies
Complies
FR