8 Parts counting
With parts counting you can either count parts into a container or remove parts from
a container. To count a greater number of parts the average weight per part has to
be determined with a small quantity (reference quantity). The larger the reference
quantity, the higher the counting exactness. High reference must be selected for
small parts or parts with considerably different sizes.
8.1 Determination of the reference weight by weighing
Set balance to zero and tare, as required.
Place a known number of parts on the balance as reference weight
If the "Weight" display is stable, enter the number of parts via number keys.
The input is shown in the unit-weight window.
Weight
While the "Number of pieces" display is flashing (3 sec) confirm by
Dead stop control is carried out and the calculated reference weight appears on the
display.
Weight
Now you can place the parts to be counted onto the weighing plate. All quantity
parameters of your goods to be weighed are displayed:
CXB-BA-e-2124
Piece weight
Reference weight
Quantity
button.
Quantity
27