PAJ-N Series
4.2 Calibration Menu [.CAL.]
InCal or Span calibration should be performed daily and when the room temperature changes.
InCal [InCal] calibrates the balance using an internal mass. (InCal models only)
Span Calibration [SpaN] uses two weight values: zero and a weight between 50% and 100% of the capacity
of the balance.
Linearity Calibration [LIN] uses three weight values: zero, 50% of the capacity and full capacity. Generally this
calibration is not required unless testing shows thet the linearity error exceeds the linearity tolerance in the
specification table. (Not available in InCal models.)
Calibration Adjust [CAL Adj] allows adjustment to the result of the internal calibration by ± 99 divisions.
(InCal models only)
4.3 Setup Menu [.setup.]
Filter [filter] - Use the low setting (SET LO) when environmental disturbances are not present. Use the
medium setting (SET MED) for normal environments. Use the high setting (SET HI) when vibrations or air
currents are present.
Auto Tare [A-tare] - The initial item placed on the balance is assumed to be a container so it is zeroed out.
The next item is then weighed. When the pan is cleared the balance resets, waiting for a container. (SET OFF,
SET ON). Local weights and measures authorities may require that Auto Tare be disabled.
Global Reset [Gbl rst] - Resets all settings to factory default values. (RESET)
4.4 Mode Menu [.mMode.]
Count Mode [Count] - [SET ON, SET OFF]
Average Piece Weight (APW) Optimization [aPwW.OPT] - (SET ON, SET OFF)
4.5 Unit Menu [.Unit.]
The Unit menu is used to enable or disable a specific unit. (SET ON, SET OFF) The unit is indicated by a small
character next to Unit in the display (g = grams).
4.6 Print Menu [.Print.]
Auto Print [A.Print] - Data will be continuously sent when [Continu] is set. Interval [Inter] sends data
every 1 to 3600 seconds. When Stable [Stable] will send data when the balance detects a new stable
reading. This can be a weight value only [LOad] or it can also include a stable zero [L+zero]. [OFF] disables
automatic printing.
Content [CONteNt] - The content in the data transmission can be modified. Each of the following settings can
be set on or off. Number Only [numMber] will only send the numeric result. Balance ID [Bal. ID] will add the
Balance serial number for traceability purposes. Reference [refer] will add reference information relevant to
the current mode. GLP [GLP] will send additional items to allow proper documentation of laboratory results.
Line Format [LinE Fo] - Single line format [single] will put all the data in one line separating each output
with a comma (,). Multi line format [mMulti] will put each data output on a new line. Multi +4 [mM-4LF] will
add 4 line spaces between each output.