Errors occurring during self-test
Self Test
Error 1xxx
See instruction manual
Error code
1000
1010
1020
1040
1050
1030
1060
1070
1080
1090
1020
1150
1160
1170
1200
PLT unit 997960
back
Cause
Pressure sensor defective
Pump running continuously Send device in for repair.
Pumping out taking too
long
Valve defective
Aeration through aeration
channel taking too long
Device leaking
Aeration through adapter
taking too long
Electronics fault
Remedy
Send device in for repair.
Check whether cap is securely inserted and
adapter is screwed in tightly. If so, then pump is
defective. Send device in for repair.
Send device in for repair.
Send device in for repair.
Check cap, adapter seating, and adapter filter,
and replace if necessary. Repeat self-test. If er-
rors cannot be resolved, send the device in for
repair.
Remove cap if not already done.
If adapter filter is contaminated, replace filter.
Send device in for repair.
10 Troubleshooting
79
Operating manual