Télécharger Imprimer la page

BRAND PLT Mode D'emploi page 54

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 89
4 The first steps
lasts about 12 minutes. It includes the additional differentiated testing of the internal tanks and
should be performed on a weekly basis.
What to do
1. Call up the Main
Menu
2. Set up the self-
test
3. Confirming the
self-test
4. Select the Quick
Test
5. Confirm the
Quick Test
6. Seal the system Press the supplied cap firmly
7. Start the Quick
Test
54
Operating manual
How to do it
Press or turn the select knob
to call up the Main Menu.
Turn the select knob to the
"Self-Test" setting.
Press the "Execute" button.
The "Self-Test" submenu is
called up. The last test per-
formed is shown with a frame
around it.
Turn the select knob to the
"Quick Test" setting.
Press the "Start/Execute" but-
ton.
The start menu is called up.
into the adapter (for the mul-
tichannel adapter, seal all four
channels).
Press the "Start/Execute" but-
ton.
The five components – sensor,
pump, valve, filter, and mea-
surement chamber – are
tested in succession.
Button to use
Display
Main menu
Pipette type
Self-test
Device OFF
Settings
Selection
back
Main menu
Pipette type
Self-test
Device OFF
Settings
Selection
back
Main menu
Quick Test
Full test
Selection
back
Self Test
Quick test
Full test
select
back
Self-test
Please
insert the blanking plug
Selection
back
Self-test
-0.10 hPa 1:sensor
back
PLT unit 997960

Publicité

loading

Ce manuel est également adapté pour:

703970