Technical Specifications
Output Specifications for Electrical Muscle Stimulation (EMS):
Waveform
Shape
Maximum Output Voltage (+/-15%)
@500Ω
Pulse Duration (+/-10%)
Net Charge @ 500Ω
Maximum Power Density @ 500Ω
ON Time (+/- 10%)
OFF Time (+/- 10%):
Output Specifications for Transcutaneous Electrical Nerve Stimulation (TENS):
Waveform
Shape
Maximum Output Voltage:
Pulse Duration:
Net Charge (per pulse @ 500Ω)
Maximum Average Power Density (@ 500Ω)
The values of PULSE DURATIONS, PULSE repetition frequencies and amplitudes, including any d.c. component, shall not
deviate by more than ± 20 % when measured with a load resistance within the range specified.
If confirmation is required that the Revitive works within its Essential Performance after a certain period of time, contact
the manufacturer
28
7376_IFU01_19000014.indd 28-29
7376_IFU01_19000014.indd 28-29
Biphasic
Square symmetrical with polarity reversal
Foot: 35Vp
Body: 19Vp
Foot: 450-970µs
Body: 450µs
[0.001]mC
Foot: 0.414 mW/cm²
Body: 1.52 mW/cm²
Foot: 1.90 - 8.30s
Body: 1.90 - 6.50s
Foot: 1.00 - 1.50s
Body: 1.00s
Biphasic
Rectangular
@
50 V
500Ω
100µs
0μC
4.19 mW/cm
2
Symbols
Complies with European Medical
Devices Directive (93/42/EEC)
UK Conformity Assessed
Product conforms to all applicable U.K. legislative
requirements.
Device serial number including
year (YYYY) and month (MM) of
manufacture can be found on the
#YYYYMMXXXXX
box and back of unit
Item number
Contraindications, Warnings and
Cautions
Make sure you understand these before using
Revitive
Power
Time Remaining
Intensity Level
EMS Mode
TENS Mode
Battery Indicator
Center Positive Polarity
Class II medical electrical
equipment double insulated
Type BF medical electrical
equipment
Legal manufacturer of the device
EU/EC European Authorised
Representative
Consult instructions for use
Ingress Protection Rating
EN
FCC mark
Certification mark employed on electronic
products manufactured or sold in the United
States which certifies that the electromagnetic
interference from Revitive is under limits approved
by the Federal Communications Commission.
Actegy Ltd complies with all applicable FCC rules.
The Waste Electrical and Electronic
Equipment Directive (WEEE
Directive).
At the end of the product lifecycle, do not throw
this product into normal household garbage, but
take it to a collection point for the recycling of
electronic equipment
Use-by date
YYYY MM DD
90%
Humidity, temperature and air
pressure limit for storage
20%
and transport
70°C
-20°C
1060 hPa
500 hPa
Humidity, temperature and air
75%
pressure limit for operating
30%
conditions
40°C
After any exposure to hot or cold temperatures
outside the specified operating range of
10°C
10 - 40°C allow the product to re-adjust to the
recommended operating temperatures to ensure
1060 hPa
continued product performance.
700 hPa
Indoor Use Only
Medical device does not contain
natural rubber latex
Do not disassemble
29