Centrifuges
A-1
4.1.7 Call up operating data
ATTENTION:
This should only be performed by advanced user or service engineer.
In the mode "Basic Adjustments" you can call up the operating data of the centrifuge. Please proceed as
described under point 4.1.2 to enter this program mode. Press the key "Accel/Decel" (5). In the display "Acc/Dec"
(A-2) the word "Service" flashes.
With the adjusting knob (1) the different information can be accessed:
A= previous starts of the centrifuge
H= previous operating hours S= software version
r= converter software
E= list of previous error messages
h= running time of the motor
N= Previous and remaining cycles of the installed rotor
The list of the last 99 error messages can be looked over by pressing the key "RPM | RCF" (4) and scroll through it
by the adjusting knob (1). The respective error codes appear in the display "RPM | RCF" (A-1). Please refer to
"Table 6: error messages" (see APPENDIX).
To change back to normal program mode again, switch off the centrifuge for a short period.