EN-8
ATTENTION!
Swing out rotors may be taken in operation only if all locations are filled in with either four buckets
or four carriers – do not mix buckets and carriers together!!
ATTENTION!
Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or
mechanical damage.
Do not operate with extremely corrosive substances, which could damage the rotor and buckets.
In case of any questions, please contact the manufacturer!
3.1.4 Loading and overloading of rotors
All approved rotors are listed with their maximum speed and maximum filling weight in "table 2 permissible net
weight" (See APPENDIX).
The maximum load permitted for a rotor, which is determined by the manufacturer, as well as the maximum speed
allowed for this rotor (See label on rotor), must not be exceeded. The liquids the rotors are loaded with should have
a maximum homogeneous density of 1.2 g/ml or less when the rotor is running at maximum speed.
In order to spin liquids with a higher density, the speed has to be reduced according to the following formula:
Reduced speed n
red
Example:
1,2
n
= √
red
1,7
To determine the relative centrifugal force(RCF/G-force) for a specific adapter, you can calculate per DIN 58 970
using the attached formula:
-
5
RCF = 1.117862*10
n: revolutions per minute (RPM)
r
: max centrifuging radius in cm by using the bottom of tubes
max
3.1.5 Removing the rotor
Untighten the rotor fixing nut completely (screw over the stiff point) and lift the rotor vertically out of the centrifuge.
(See figure 12 and 13).
3.2 Lid control
3.2.1 Lid open
After the run, when the lid of the centrifuge is closed, the word "close" (M1) appears in the display "RPM | RCF"
(A-1). Additionally, if there is a rotor in the centrifuge, the word "rotor"(M3) appears as well as the code number of
the respective rotor, which is in the centrifuge system "71" (M4). If there is no rotor in the centrifuge, the word
"rotor" (M3) flashes and additionally the word "no" (M4) appears. By pressing the key "Door Open" (7) you can
release the lid of the centrifuge. As soon as the electromagnetic lid is completely released, the word "open" (M2)
appears. Now you can open the lid of the centrifuge. Please refer to figure 18 below for reference.
Figure.16 WRONG
1,2
= √
x max. speed (n
ℎℎ
x 4.000 = 3.360 rpm
2
*n
*r
max
Figure.17 CORRECT
) of the rotor
max
TM
Frontier
Centrifuges