USER'S MANUAL
ADJUSTING REGULATOR AIR PRESSURE
With the Dental Unit pressurised and air lines turned ON,
the black knob on top of the regulator is pulled up from the
locked to the free position to adjust air pressure
(see picture). Depress the black knob after setting to lock
pressure adjustment.
The regulator gauge registers working or air line pressure
which may be increased by turning clockwise and lowered by
turning anti clockwise.
The pressure air line gauge will register the working pressure,
which should be 80 p.s.i. / 5BAR.
Unit starts automatically. Isolate power supply and relieve air
pressure by opening tank ON/OFF valve prior to commencing
non electrical service tasks.
This appliance is not intended for use by persons (including
children) with reduced physical, sensory or mental capabilities,
or lack of experience and knowledge, unless they have
been given supervision or instruction concerning use of the
appliance by a person responsible for their safety
High voltage used on this equipment.
Isolate from mains before removing any covers, which
should only be carried out by a qualified electrician.
Ensure power outlet is adequately rated for this unit.
FEATURE
DAILY
Silent
Hurricane
Compressor
Regulator -
Automatic, when tank
Auto Drain
drained of moisture
Filter
Check that correct
Regulator -
air pressure is being
Air Pressure
delivered
Contact iM3 for your yearly compressor service kit.
Failing to perform your annual services
WILL VOID UNIT WARRANTY
GS DELUXE / (LED) VETERINARY DENTAL UNIT
WEEKLY
Turn power off.
Drain moisture from main compressor
tank by opening the valve on the top
of the white tank marked with a yellow
sticker—Drain.
Catch the fluid from the white drain
tube into a container. Keep the drain
open until all air is removed from tank.
See Adjusting Regulator Air Pressure on previous page.
AIR TANK
REGULATOR
PRESSURE GAUGE
PRESSURE GAUGE
Fig.32 - Handpiece HS / LS Pressure
Adjustment knobs
MONTHLY
YEARLY
Replace regulator filter.
30 � EN