USER'S MANUAL
4. O P E R AT I O N
1
Turn on the compressor and
allow to pressurize.
4
Check handpiece operation:
a) Select HS on the Toggle,
Remove High Speed Handpiece
from holder and depress
handpiece foot control pedal—
handpiece should now operate.
b) Select LS and Remove Low
Speed Handpiece from holder
and depress handpiece foot
control pedal—handpiece should
now operate.
c) Adjust individual handpiece
pressure if required, (see page
26). Each handpiece—Low speed
and high speed has a specific
pressure requirement.
Please observe manufacturers handpiece
WARNING:
maintenance instructions for handpiece longevity.
GS DELUXE / (LED) VETERINARY DENTAL UNIT
2
2
PRESSURE GAUGE
Make sure your compressor is set
to deliver the correct working
pressure. The air pressure
regulator gauge should read
5.5BAR (80 psi). If adjustment
is required, please see table for
compressor pressure adjustment
instructions. The pressure is set at
the iM3 factory.
2
5
Pressurize water system by
selecting the coolant bottle
pressure switch to ON, located
on the front of the control panel.
(as pictured above)
IMPORTANT: When filling
distilled water and/or CLS
Enviro bottles de-pressurize
the water system, by
switching the coolant bottle
pressure switch to OFF.
Check water and air supply to the
3 way air water syringe. Depress
the right hand button for air and
left hand button for water or
both together for a mist.
3
REGULATOR
Place the foot control pedal on
the floor in front of the unit. The
foot pedal operates the High and
Low speed handpieces when
depressed.
3 6
Water volume. Remove high
speed handpiece, from holder
and depress foot control pedal.
Adjust water spray volume to
a fine mist by rotating the flow
control knob. (as pictured above)
17 � EN