USER'S MANUAL
iM3 DISTILLED WATER AND CLS
ENVIRO SOLUTION SUPPLY
Each iM3 dental system is supplied with an amber and
clear water bottle which should be filled only with distilled
water to assist in preventing mineral or bacteria build up in
the handpieces. This water bottle supplies the high speed
handpieces, 3 way air water syringe and ultra sonic scaler
coolant outlet.
The CLS Enviro Solution supply and water system
allows you to choose between distilled water or CLS
Solution at the flick of a switch, depending on the dental
procedure you wish to perform. The CLS solution contains
chlorhexidine which is light sensitive and should be kept in
the amber bottle provided.
The coolant system for your iM3 dental system, has been
preset at 25 psi.
REFILLING COOLANT BOTTLES
To refill the distilled water or CLS solution, flip the coolant
pressure switch (as pictured right) on your dental system to
OFF - depressurizing the system. After the system has been
depressurized, unscrew the bottle and refill with either
distilled water or CLS Solution. After filling the coolant
bottle screw it back into the holder and flip the coolant
switch to ON, repressurizing the system.
DO NOT OVERTIGHTEN THE BOTTLES.
PLEASE NOTE: When the day's dental procedures
are complete, flip the ON/OFF coolant switch
to the OFF position, depressurizing the coolant
bottles for storage.
Here's what Dr Jeanie Hawkins, DVM, Diplomat AVDC had
to say about CLS Solution and the Auto Flush System
"Having CLS Solution at my fingertips
to flush periodontal pockets is a real
plus. iM3
is the only Company to
®
address the problem of cleaning the
airlines and high speed handpieces."
GS DELUXE / (LED) VETERINARY DENTAL UNIT
CLS
NOTE: Failure to use distilled water will
void the iM3 machine warranty.
Fig.1 - Coolant Pressure ON / OFF Switch located on
Units front control panel
CLS Solution 250 ml Concentrate - single
SKU EU/AU/US: C3150
CLS Solution 250 ml Concentrate - 6 pack
SKU EU/AU/US: C3150-6
18 � EN
ONLY
Distilled
Water
WATER