7.7.3 Gravitational constant adjustment location < graadj >
TKXC-TM-A-BA_IA-def-2310
Press and hold the TARE and ON/OFF buttons simultane-
ously to enter the setup menu.
Wait until the first menu item < Cal> is displayed.
Confirm by ➔ button, < CalExt> will be displayed.
Use the navigation keys to select < graadj>.
Acknowledge using ➔ button, the current setting is dis-
played. The active digit is flashing.
Enter weight value and confirm using the ➔ button, numer-
ical input see chap. see chap. 3.2.2.
Weighing balance returns to menu.
Press repeatedly button to exit menu.