6 Error limits
6 Error limits
20 °C
!
Ex
Error limits
Nominal volume
ml
2
10
25
*A = Accuracy, CV = Coefficient of Variation
Partial volume
The percentage values for A and CV are relative to the nominal volume (V
partial volumes (V
V
N
A
=
T
V
T
e.g.
V
N
V
= 50% N
T
V
= 10% N
T
*A = Accuracy, CV = Coefficient of Variation
The maximum error for a single measurement can be calculated from the sum of error limits EL = A + 2
× CV (e.g. for 10 ml: 120 µl + 2 x 20 µl = 160 µl).
66
Operating manual
Error limits in relation to the nominal capacity (= maximum
volume) indicated on the instrument at equal temperature
(20 ºC/68 °F) of instrument, ambient environment and dis-
tilled water. Testing takes place according DIN EN ISO 8655-
6 with a completely filled instrument and with uniform and
smooth dispensing.
A* ±
%
1.2
1.2
1.2
).
p
.
A
N
Volume
10
5
1
A* ±
CV*
µl
%
24
0.2
120
0.2
300
0.2
A* ±
A* ±
%
μl
1.2
120
2.4
120
12
120
NOTICE
CV*
µl
4
20
50
) and must be converted for
N
CV*
CV*
%
μl
0.2
20
0.4
20
2.0
20