2.2.1 Operation
When the instrument is correctly used, the dispensed liquid comes into contact with only the follow-
ing chemically resistant materials:
seripettor®
FEP, PP, PE, EPDM
seripettor® pro
PP, PE, borosilicate glass, Al
2.3 Limitations of Use
This instrument is designed for dispensing liquids, observing the following physical limits:
Use between +15°C and +40°C of instrument and reagent (seripettor®: agar culture medium up to
max. 60 °C)
Vapor pressure up to 500 mbar
Density up to 2.2 g/cm
Kinematic viscosity*:
2 ml instrument: 300 mm
10 ml instrument: 150 mm
25 ml instrument: 75 mm
* dynamic viscosity [mPas] = kinematic viscosity [mm
2.4 Operating Limitations
Liquids, which form deposits may make the piston difficult to move or may cause jamming (e.g.,
crystallizing solutions or concentrated alkaline solutions). If the piston movement becomes slug-
gish or stiff, the instrument should be cleaned immediately (Cleaning, p. 68).
When dispensing inflammable media, make sure to avoid to buildup of static charge, e.g., do not
dispense into plastic vessels; do not wipe instruments with a dry cloth.
The instrument is designed for general laboratory applications. Compatibility of the instrument
for a specific application (e.g., trace material analysis, food sector etc.) must be checked by the
997468
-O
-ceramic, ETFE, PFA, FEP, PTFE, Pt-Ir
2
3
3
2
/s
2
/s
2
/s
2
3
/s] x density [g/cm
]
2 Safety Instructions
51
Operating manual