5.9.1 Heart Rate Measurement Distance method
"Left" click mouse to activate the cursor "+". Place the "+" at the start position of the
measurement, "Right" click mouse to lock the cursor in place. Place the second "+" at the end
point of the measurement (repeatedly "Right" click mouse to switch between the start and
ending points). "Left" click mouse to get distance value. Repeat the above described method
to obtain additional measurements.
5.9.2 Heart Rate Measurement Circumference and area method
Click the "Left" mouse button to activate the cursor "+". Move the "+" to the beginning of the
area that needs to be measured, then Click the "middle" mouse button to lock the cursor in
place at the starting point. Move the "+" to circle over the area that needs to be measured.
"Middle" click mouse to get the value. Repeat the process to obtain additional
measurements.
Note: In circumference and area measurement, when the trace is not totally closed, the
starting point and the ending points are automatically connected to form a closed
measurement trace.
5.10 Data input
To input data, "freeze frame" image, press NOTE. Shift cursor to "ID" and enter data – limit 6
numeral characters. After entering ID, the cursor will move to "AGE", enter data – limit 3
numeral characters. After entering age, the cursor will move to "SEX". Press M/0 to input
female symbol, or press B M/9 to input male symbol.
Pressing NOTE again will erase all input data.
5.11 photo shots + image printing
• Photo Shots
Freeze frame image, then use Polaroid camera or digital camera to take photos.
• Image Printing
Connect "video out" of MS Multiscan to the video input on printer (connection cable sold
Separately). Switch on power supply and operate printer in accordance with the
manufacturer's instructions.
5.12 Power Off
Turn off the power supply switch. Disconnect the AC-adaptor if the scanner is
not being used for a long time.
Note: Do not disconnect or re-connect the power supply until the scanner has shut down.
Wait 2 or 3 minutes before restarting the scanner.