SECTION 5—TYPICAL PRODUCT
PARAMETERS
IPX1
Electrical
Requirements:
Rated Current Input:
Sound Level:
Altitude:
Oxygen Output
Concentration
Levels:
Maximum Outlet
Pressure:
Part No 1163145
SECTION 5—TYPICAL PRODUCT PARAMETERS
Alternating Current
Type BF equipment
Unit running
Unit not running
Attention - Consider Accompanying Documents
DO NOT smoke
Class II, Double Insulated
Protected against dripping water
230 VAC ± 10% (253 VAC/207 VAC), 50 Hz
1.4 A
39.5 dBA Average
All Perfecto
Models: Up to 8,000 ft (2438 meters) above sea
2
level without degradation of concentration levels. From 8,000 ft to
13,129 (4000 meters) below 90% efficiency.
All Perfecto
Models while filling a HomeFill System:
2
Up to 6,000 ft (1828 meters) above sea level without degradation
of concentration levels. From 6,000 ft (1828 meters) to 13,129 ft
(4000 meters) below 90% efficiency.
87% to 95.6% at 0.5 to 5 L/min
93% ± 3% at 2 L/min
NOTE: Stated concentration levels achieved after initial
warm‐up period (approximately 30 minutes)
34.5 kPa ± 3.45 kPa (5 psi ± 0.5 psi)
13
Perfecto
™Series
2