6.4 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1).
For this warm-up period the scale must be connected to the power supply (batteries).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
6.5 Type of protection IP65
The KERN WTB meets the requirements for IP65 protection type.
Suitable for a short-time contact with liquids. For cleaning use damp cloth. Dust-tight.
7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
7.1 Adjustment
Carry out adjustment as close as possible to the maximum load of the scales )see
chap. 1 „Technical data"). The accuracy of the adjustment weight must correspond
approximately to or, if possible, be better than, the readability d of the balance.
Information about test weights you will find in the internet under
http://www.kern-
sohn.com
Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
WTB-N_WTB-NM-BA-e-1823
20