7.
Open the meter lid. Insert the tube
into the chamber. Align the index line
on the tube with the index arrow on
the meter. Close the lid.
8.
Press
ENTER
and scan the sample. The screen will
display Scan Sample and Sample
Done for about 1 second. The result
will appear on the screen.
Press
Print. Press ENTER to print to the BLE
Mobile Printer (Code 5-0067).
ANALYSIS WITH BLANKING
To obtain the most accurate results the meter should be blanked before measuring a
sample. The blanking step is not as critical for samples above 10 NTU. The meter should
always be blanked before reading samples below 10 NTU.
1.
Press and briefly hold
the meter on. The LaMotte logo screen
will appear for about 3 seconds and
the Main Menu will appear.
2.
Press
ENTER
3.
Press
With Blank.
to select Scan Sample
or
to scroll to
to turn
to select Measure Menu.
to scroll to Turbidity -
Turbidity NB (N)
10.22
Scan Blank
Scan Sample
12:00:00 p
Main Menu
Measure Menu
Setup Menu
12:00:00 p
Measure Menu (N)
Turbidity - No Blank
Turbidity - With Blank
12:00:00 p
Measure Menu (N)
Turbidity - No Blank
Turbidity - With Blank
12:00:00 p
39
NTU
1/500
1/500
1/500
1/500