LaMotte 2020t BLE Mode D'emploi page 17

Publicité

Les langues disponibles

Les langues disponibles

8.
Press
desired option.
9.
Press
ENTER
The screen will display Storing... for
about 1 second and return to the
Turbidity Options menu. Press
to return to a previous menu.
NOTE: When the Averaging option is enabled, more time will be required to display a reading and more
power will be used.
„ AUTO CALIBRATION
Auto Calibration will calibrate the meter to 1, 10 or 100 NTUs or FTUs depending on which
Calibration Curve is chosen. The Auto Calibration feature can be used for environmental
testing of natural waters or samples with moderate amounts of turbidity when less
precision is required. Auto Calibration is not as accurate as a manual calibration. It should
not be used for compliance monitoring. Follow the manual Calibration procedure for
compliance testing.
„ SETTING THE CLOCK
1.
Press and briefly hold
the meter on. The LaMotte logo screen
will appear for about 3 seconds and
the Main Menu will appear.
2.
Press
3.
Press
ENTER
Press
or
to scroll to the
to save the selection.
EXIT
to turn
to scroll to Setup Menu.
to select Setup Menu.
to scroll to Set Clock.
Disabled
2 Measurements
5 Measurements
10 Measurements
12:00:00 p
Turbidity Options
Turbidity Calibration
Turbidity Units
Averaging
12:00:00 p
Main Menu
Measure Menu
Setup Menu
12:00:00 p
Setup Menu
Measure Menu
Setup Menu
12:00:00 p
Setup Menu
Turbidity Options
Auto Calibration
Set Clock
Logging
12:00:00 p
17
Averaging
1/500
1/500
1/500
1/500
1/500

Publicité

loading

Ce manuel est également adapté pour:

2020i ble