LaMotte 2020t BLE Mode D'emploi page 34

Publicité

Les langues disponibles

Les langues disponibles

3.
Press
ENTER
4.
Press
ENTER
Options.
5.
Press
ENTER
Calibration.
6.
Scroll to the desired calibration option.
NOTE: Stablcal
standards below 50 NTU should not be used to calibrate the 2020t/i BLE. The diluent
®
has a different refractive index than traditional formazin standards and will affect the results.
7.
Press
ENTER
The screen will display Storing... for
about 1 second and return to the
Turbidity Options menu.
to select Setup Menu.
to select Turbidity
to select Turbidity
to save the selection.
Setup Menu
Turbidity Options
Auto Calibration
Set Clock
Logging
12:00:00 p
Turbidity Options
Turbidity Calibration
Turbidity Units
Averaging
12:00:00 p
Turbidity Calibration
Nephelometric NTU
Ratiometric NTRU
Attenuation AU
12:00:00 p
Turbidity Calibration
Nephelometric NTU
Ratiometric NTRU
Attenuation AU
12:00:00 p
Turbidity Options
Turbidity Calibration
Turbidity Units
Averaging
12:00:00 p
34
1/500
1/500
1/500
1/500
1/500