3.
Press
ENTER
4.
Press
ENTER
Options.
5.
Press
ENTER
Calibration.
6.
Scroll to the desired calibration option.
NOTE: Stablcal
standards below 50 NTU should not be used to calibrate the 2020t/i BLE. The diluent
®
has a different refractive index than traditional formazin standards and will affect the results.
7.
Press
ENTER
The screen will display Storing... for
about 1 second and return to the
Turbidity Options menu. Press
to return to a previous menu.
SELECTING TURBIDITY UNITS
The results for each calibration curve will be reported in the default units for that calibration
curve unless different units are chosen. The default unit is shown after the calibration curve
name. The default units are:
to select Setup Menu.
to select Turbidity
to select Turbidity
to save the selection.
EXIT
Setup Menu
Turbidity Options
Auto Calibration
Set Clock
Logging
12:00:00 p
Turbidity Options
Turbidity Calibration
Turbidity Units
Averaging
12:00:00 p
Turbidity Calibration
Nephelometric NTU
Ratiometric NTRU
Attenuation AU
12:00:00 p
Turbidity Calibration
Nephelometric NTU
Ratiometric NTRU
Attenuation AU
12:00:00 p
Turbidity Options
Turbidity Calibration
Turbidity Units
Averaging
12:00:00 p
13
1/500
1/500
1/500
1/500
1/500