Molift Air 500 / BM16099 Rev F / www.etac.com
Hand control
The Hand control has 2 buttons for lifting and lowering, or 4 but-
tons if the hoist is equipped with propulsion. The Hand control
has an indicator light that will illuminate when battery level is low
and the lifter requires charging
Do not pull Hand control to move the hoist
along the rail
The lift will not work if hand control is
connected to a charger.
Disconnect hand control from charger before
operating the lift.
Mounting sling bar
2
1
3
4
1.
Align sling bar in connection point.
2.
Push button on locking pin and insert all the way through.
3.
Make sure locking pin is properly fastened
4.
Push and hold down the button on the locking pin, and pull
to remove locking pin.
Emergency stop / emergency lowering
Emergency Stop
Activation and reset of emergency stop:
Pull to activate emer-
gency stop. The button
will come out, and hoist
will stop.
Push with finger or use tube
on cord to push button
back in to reset emergency
stop.
Electrical Emergency Lowering
Pull and hold to start lowering. Hold until client is
lowered and can be released from sling bar.
Only for emergency use when hand control
does not work!
Manual Emergency lowering
Allen key or bit for power drill can be found inside the lifter.
Remove covers on both sides of the hoist to locate.
To open side covers use a screwdriver or power drill on all 8
screws (4 on each side). Turn the red "Brake Release" handle
up. Locate the tools on the other side of the hoist and rotate
counterclockwise to lower lifting belt. Using a power drill is
recommended.
Service is NOT required if:
1) the Air manual emergency down is used for a known reason,
and
2) the belt is undamaged, and
3) the unit is operating normally after the manual emergency
lowering procedure.
The new brake override will contribute to this by making emer-
gency manual lowering less stressful to lift components.
If the unit have had a malfunction with unknown reason, and
manual emergency lowering system has been used, Service is
required.