4. Confirm the weighing time by pressing Yes, and the
message "Ready" will display at the lower left of the
screen.
5. Place the dynamic object on the pan. The balance
begins a countdown (averaging process). During the
countdown, the screen shows the time remaining.
6. When the countdown ends, the result line is displayed
and held.
7. After the dynamic object is removed, the weight will
be automatically set to zero, and the balance will
return to the status of "Ready".
Item Settings
1. Averaging Time: Set the averaging time to a value between 1 and 15 seconds. Default is 10 seconds.
2. Printing Setup: Changing printing setup. See Section 7 for more information.
4.5
Density Determination
Note: Before using any application, be sure the balance has been leveled and calibrated.
Use this application to determine an object's density.
A Density Determination Kit, Part Number 80253384, is designed to be used with PX series balances.
Illustrations in this procedure refer to the density kit, however, you may use whatever lab apparatus that will suit
the requirements for density measurements. A built in reference density table for water at temperatures between
10ºC and 30.9ºC is included in the balance software. Review this entire section before attempting density
measurements.
Operation Method
Press and hold the Function / Mode button until the Density appears on the screen.
After confirmation by pressing Yes, the message "Change parameter?" will display on the screen. The settings
can be kept or changed as desired.
Item Settings:
Sample Type: Solid, Liquid
Auxiliary Liquid: Water, Alcohol, Other
Porous Material: Off, On
Water Temperature: 20
Alcohol Temperature: 20
Volume (of Calibrated Sinker): 10 ml (by default)
Weight (of Porous Material): 5.000 g
(by default)
(by default)
EN-13