control.
address
Controller address
Information obtained by the serial interface are listed in the
following table where the column "TYPE" points out a logic variable
with "D" (allowed values "000000" or "111111") and a numeric
variable with "N".
Variables with the first character of "Parameter Number" equal to X
can be either written or read. Variables with that character equal to
"0" can be only read, equal to "1" can be only written.
Tab. 2
PARAMETER
NUMBER
X0002
X0003
X0008
10009
00302
00303
Turbo-V 81 PCB 24 Vdc Controller User Manual / 87-900-121-01 (D)
in the case of an attempt of writing a logic parameter with a value
not equal to " 000000" or " 111111", or when the parameter is a
read only one, with:.
1
0
param.
address
in the case of a time between two characters greater than 1
second, with:
TYPE
D
D
D
D
D
N
Technical Information
Data Exchange Protocols
0
6
-
L
O
G
N
A
DESCRIPTION
Low speed ON/OFF
Pump ON/OFF
Remote ON/OFF
Error acknowledgment: error
status clearing
K1 status
Error code: error code
displaying (see the following
table)
17
I
C
check-
CR
sum
K
CR
189/208