Use
This section provides all the major operating procedures. For further
details and for the procedures that involve connections or specific
options, refer to the section entitled "USE" in the Appendix
"Technical Information". Perform all electrical and pneumatic
connections before using the controller, with reference to the manual
of the connected pump.
WARNING!
If the pump is placed on a table, make sure that it is stable so as to avoid
injury to persons and damage to the equipment. Never operate the pump
unless the input flange is connected to the system or closed using the closure
flange.
Operating Procedures
Powering on the Controller
To power on the controller, simply provide electrical power to it.
Starting the Pump
To start the pump, leave pin 12 of connector J4 open so that the
pump starts-up when 24 Vdc is provided to the controller.
Stopping the Pump
To stop the pump, short-circuit pins 12 and 15 of connector J4.
Turbo-V 81 PCB 24 Vdc Controller User Manual / 87-900-121-01 (D)
Instructions for Use
Use
169/208