Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6. Faults

6. Faults
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the instrument
must be taken out of operation immediately.
EN
WARNING!
Physical injuries and damage to property and the environment caused
by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable or
toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioac-
tive), and also with refrigeration plants and compressors, there is a danger of
physical injuries and damage to property and the environment.
Should a failure occur, aggressive media with extremely high temperature
and under high pressure or vacuum may be present at the instrument.
6.1 Reasons and measures
Fault
Indication error
Window misted up on inside
Functionality of the 'as-deliv-
ered' condition is no longer
fulfilled
12
Ensure that pressure or signal is no longer present and protect against
accidental commissioning.
Contact the manufacturer.
If a return is needed, please follow the instructions given in chapter 9.2
"Return".
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.
Wear the requisite protective equipment (see chapter 3.6 "Personal
protective equipment").
Cause
Inadmissible operating condi-
tions, e.g. exceeding the
pressure limitation
Pressure compensation film
on the window has come
loose
Transport damage
WIKA operating instructions models 116.15 and 116.18
Measure
Replace pressure gauge , see
chapter 8 "Dismounting, return
and disposal"
Replace pressure gauge , see
chapter 8 "Dismounting, return
and disposal"
Return to the manufacturer

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

116.18

Table des Matières