6.2 Sub total and total counter

The grease meter is equipped with a SUB TOTAL and TOTAL counter that can be reset.
Go to the counter option in the menu (see 6.1) to display this value and possibly reset it.
The display shows:
Press ENTER (the left button).
The display shows:
This shows you the SUB TOTAL amount of grease
that has passed the meter since the last reset.
To reset the sub total counter, press ZERO (right button).
To exit the menu press EXIT (left button).
The TOTAL counter is set in the same manner.

6.3 Calibration of the meter

The grease meter can be calibrated for different grease types, characteristics and densities.

6.3.1 Calculate the correction factor

Correction factor = real quantity / shown quantity
Example:
500 grams of grease have been dispensed.
The grease meter shows 485 GRAM.
Correction factor = 500 / 485 = 1,03

6.3.2 Calculate the calibration factor

The new calibration factor = the old calibration factor × the correction factor.
Example:
The old calibration factor is 700.
The correction factor is calculated as 1,03.
The new calibration factor = 700 × 1,03 = 721
SKF LAGM 1000E
7