MicroLim
Line Isolation Monitor
Section 5—Operation
Factory Default Settings
The dip switch, jumper, and slide switch on the LIM are factory-set as shown in table 3.
For LIMs used in Canada (catalog no. MLIM), alarm level must be set to 2 mA to meet CSA
certification.
No.
1
2
3
4
5
1
1
Normal Operation of Front Panel
The LIM front panel (operator interface board) was designed for maximum ease of use.
Figure 3 on page 7 shows the LIM front panel. The following paragraphs describe
normal operation of the LIM front panel.
Normal Start-Up Sequence
The normal start-up sequence is as follows:
1.
The Total Hazard Current display counts down: HI, 99, 88, ..., 11, 00.
2.
The Analog Bar Graph rapidly increases from no bars illuminated, to all bars illu-
minated. The red bars blink approximately every second.
3.
The Safe LED is illuminated.
4.
The Hazard LED is illuminated.
5.
Total Hazard Current displays "CL."
6.
Total Hazard Current displays "CH."
7.
Total Hazard Current displays the actual predicted hazard current. Start-up
sequence is complete.
Table 3 Factory Default Settings
Possible Settings
Dip Switch Settings
Quiet—Loud
5 mA—2 mA (2 mA Canada)
B—A
OFF (Mute)—ON (Alarm)
Single Phase — Three Phase
Jumper Setting
LO LINE — HI LINE
Slide Switch Setting
120V—240V
© 1996 Federal Pioneer All Rights Reserved
Bulletin No. 43134-017-02
Factory Default Settings
Loud
5 mA
B
Alarm ON
Single Phase
HI LINE
240V
Page 11
8/96
E
N
G
L
I
S