Negative Check
Negative check is used to determine when the material removed from the balance is within the
target range. In this case the UNDER and OVER limits are both negative values.
(The UNDER limit must be greater than the OVER limit.)
Place the item to be weighed on the balance and press TARE.
Remove a portion of the item until it is within the ACCEPT range.
Zero Check
Zero check is used when comparing subsequent samples to an initial reference sample. In this
case, the UNDER limit must be a negative value and the OVER limit must be a positive value.
Place the reference item on the balance and press TARE. Remove the reference sample and
place the item to be compared on the balance to determine if it is within the ACCEPT range.

4.7 Display Hold

The Display hold application allows the user to capture and store the highest stable weight in a
series of weight measurements.
In the upper left portion of the home screen, select Display hold.
The DISPLAY HOLD Home screen:
4.7.1 Item Settings
To view or adjust the current settings:
Touch the Item Settings button. The Settings
screen appears.
Mode: Select the mode to either Peak Hold or
Display Hold (default).
Sample Name: Enter a sample name.
Weighing Units: Select the displayed unit.
Print Settings: To view or adjust the print
settings