8 Scope of supply
Fault
Cause
Mask is not
adjusted
correctly.
Mask
cushion is
damaged.
Therapy
Tube system
pressure is not
is leaking.
reached.
Air is escap-
ing from
pressure
measure-
ment con-
nection.
8 Scope of supply
Size
Variant
S
vented
M
vented
vented
L
You can order replacement parts separately
if necessary. A current list of replacement
parts can be ordered on the Internet at
www.weinmann.de or via your specialist
dealer.
18
EN
JOYCEeasy
All manuals and user guides at all-guides.com
Remedy
Re-adjust mask
(see
4.2,
p. 15).
Replace mask
cushion.
Check connection
and fit of tubes.
Close pressure
measurement port
with plug (included in
scope of supply).
Item number
WM 25510
WM 25520
WM 25530
9 Technical data
Product class according to
directive 93/42/EEC
Dimensions (W x H x D)
Weight
Therapy pressure range
Hose connection: Cone
according to EN ISO 5356-1
Pressure measurement port
Temperature range:
Operation
Storage
Flow resistance
at 50 l/min
at 100 l/min
Quoted two-figure noise
emission value according to
ISO 4871:
-Sound pressure level
-Sound power level,
uncertainty factor:
Service life
Standards applied
1
depending on size and variant.
2
The materials of the mask deteriorate if exposed to
e.g. aggressive detergents. In individual cases,
therefore, it may be necessary to replace mask parts
sooner.
The right to make design modifications is
reserved.
Nasal cannula
IIa
approx. 80 mm x
1
130 mm x 90 mm
1
ca.72 g
4 hPa - 30 hPa
Ø 22 mm (male)
Ø 4 mm
+5 °C to + 40 °C
-20 °C to +70 °C
0.05 hPa
0.28 hPa
18 dB(A)
23 dB(A)
3 dB(A)
2
up to 12 months
EN ISO 17510-
2:2009

Publicité