14.1 Adjustment
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Display
z000
��
a
SETUP
��
UNKT
��
B
CAL
��
K
CAL U
��
MCB_MPT-BAPS-e-1710
Operation
Start balance by pressing
Press
for approx. 3 sec,
in the display appears „SETUP", followed by „UNIT"
TARE
Press
repeatedly until "CAL iB" appears
HOLD
Press
, „CAL U" appears
TARE
, the appeared triangle must be located in the
Press
upper left part of the display.
If not, press again
ON
OFF
TARE
.
39