Technical Specifications
.
Operating pressure
range
Dimensions
(H x W x D)
Weight
Power Supply
Air Filter
Air Tubing
Air Outlet
IEC 60601-1
Classifications
Noise Level
Supplemental Oxygen
Housing Construction
Operating
Temperature
Operating Humidity
Storage and Transport
Temperature
Storage and Transport
Humidity
Operating Altitude
Electromagnetic
Compatibility
Note: The manufacturer reserves the right to change these specifications without notice.
Symbols which appear on the device
Attention, consult accompanying documents;
Dangerous voltage;
All manuals and user guides at all-guides.com
4 to 20 cm H
O
2
4.6" x 6.5" x 5.7"
(112 mm x 164 mm x 145 mm)
2.9lb (1.3 kg)
Input range for S8 Escape II with HumidAire 3i:100–240V, 50/60Hz;
110V, 400Hz; 2.5A < 140VA (110W) (maximum power consumption).
Instantaneous peak power consumption <340VA.
Refer to the DC-12 converter instructions for DC ratings. Only use
the ResMed DC-12 converter for DC input.
Typical power consumption at 20 cm H
Two-layered, powder-bonded, polyester non-woven fiber
Flexible plastic, 6'6" (2 m)
The 22 mm conical outlet complies with EN 1281-1
Class II (double insulation), Type CF
Radiated sound pressure is measured at <30dBA at 10 cm H
3'3.36" (1 m) according to ISO17510-1.
Recommended maximum supplemental oxygen flow: 4 L/min
Flame retardant engineering thermoplastic
+41ºF to +95ºF (+5ºC to +35ºC)
10%–95% non-condensing
-4ºF to +140ºF (-20ºC to +60ºC)
10%–95% non-condensing
Sea level to 8500' (2591 m)
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments.
The electromagnetic compatibility tables for these ResMed devices can
be found on www.resmed.com, on the Products page under Service
& Support. Click on the PDF file for your device.
Class II equipment;