Télécharger Imprimer la page

ResMed S8 Escape II Guide Utilisateur page 16

Masquer les pouces Voir aussi pour S8 Escape II:

Publicité

Les langues disponibles

Les langues disponibles

Insufficient air delivered from S8 Escape II
Ramp Time is in use.
Air filter is dirty.
Air tubing is kinked or
punctured.
Air tubing is not connected
properly.
Mask and headgear are not
positioned correctly.
Cap(s) are missing from
access port on mask.
Pressure required for
treatment may have changed.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
Data Card is not inserted
correctly.
The Data Card is only partly
inserted.
You may have removed the
Data Card before settings
were copied to the
S8 Escape II.
The card is not a ResScan Data
Card.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
The identification details on
the Data Card do not match
the identification details on
your device.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
There is a data error on the
Data Card.
The following message is NOT displayed on the LCD after you try to update the
settings using the Data Card: Settings Success Remove Card
The settings were not
updated.
12
All manuals and user guides at all-guides.com
Wait for air pressure to build up or change ramp time.
Replace air filter.
Straighten or replace tubing.
Check air tubing.
Adjust position of mask and headgear.
Replace cap(s).
See your clinician to adjust the pressure.
Card Error Remove Card
Check that the patient is inserting the Data Card into the
Data Card module with the arrow facing up, while the S8
Escape II is switched on and not delivering therapy
(standby mode).
Ensure that the Data Card is inserted into the Data Card
slot as far as it will go.
Reinsert the Data Card and wait for the Settings
Success Remove Card message to appear on the
LCD.
Only use a ResScan Data Card.
Settings Invalid Remove Card
Contact your clinician immediately.
Settings Error Remove Card
Contact your clinician immediately.
Contact your clinician immediately.

Publicité

loading