PROCOM 2 or CYCLIC 2020 regulation - SEC 7HC air treatment type
15 ELECTROMAGNETIC COMPATIBILITY
The air unit with PROCOM2 or CYCLIC 2020 automate needs special precautions
regarding EMC. The air unit with PROCOM2 or CYCLIC 2020 automate has to be
installed and put into service according to the EMC requirements joined in the following
tables.
519015-15
HOSPITAIR CABIN (MV) – 11 & 15 bar
Guidance and manufacturer's declaration – electromagnetic emissions – for all ME equipment and ME systems
The air units with PROCOM2 or CYCLIC 2020 automate is intended for use in the electromagnetic environment specified below. The
customer or the user of the air units with PROCOM2 or CYCLIC 2020 automate should assure that it is used in such an
electromagnetic environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Complies
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity - for all ME equipment and ME systems
The air units with PROCOM2 or CYCLIC 2020 automate is intended for use in the electromagnetic environment specified below. The
customer or the user of the air units with PROCOM2 or CYCLIC 2020 automate should assure that it is used in such an
electromagnetic environment.
Immunity test
IEC 60601 test level
Compliance level
For 230 Vac / 50Hz
Electrostatic
discharge (ESD)
±6 kV contact
±6 kV contact
IEC 61000-4-2
±8 kV air
±8 kV air
For 230 Vac / 50Hz
Electrical fast
±2 kV for power supply
2 kV for power supply
transient/burst
lines
IEC 61000-4-4
±1 kV for input/output
±1 kV for input/output
lines
For 230 Vac / 50Hz
Surge
±1 kV line(s) to line(s)
±1 kV mode differential
IEC 61000-4-5
±2 kV line(s) to earth
±2 kV mode common
<5 % UT (>95 % dip in
Voltage dips, short
UT) for 0,5 cycle
interruptions and
40 % UT (60 % dip in UT)
Pour 230 Vac / 50 Hz
voltage variations on
for 5 cycles
0 Vac during 10 ms
power supply input
70 % UT (30 % dip in UT)
92 Vac during 100 ms
lines
for 25 cycles
161 Vac during 1 s
IEC 61000-4-11
<5 % UT (>95 % dip in
0 Vac during 5 s
UT) for 5 s
Power frequency
(50/60 Hz)
3 A/m
3 A/m à 50Hz
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Electromagnetic environment – guidance
The air units with PROCOM2 or CYCLIC 2020 automate uses RF
energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby
electronic equipment.
The air units with PROCOM2 or CYCLIC 2020 automate is suitable
for use in all establishments, including domestic establishments and
those directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes.
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of a typical
lines
commercial or hospital environment.
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
air units with PROCOM2 or CYCLIC 2020 automate
requires continued operation during power mains
interruptions, it is recommended that the air units with
PROCOM2 or CYCLIC 2020 automate be powered
from an uninterruptible power supply or a battery
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
No components sensible in product to magnetic field.
31