Télécharger Imprimer la page

Seca 634 Mode D'emploi Et Garantie page 84

Masquer les pouces Voir aussi pour 634:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 101
Setting filtering (Fil)
Switching unit of weight
84 •
Filtering (Fil) allows you to reduce interference (e.g. as a
result of patient movement) when determining weight.
1. Select the item "Fil" from the menu.
2. Confirm your selection.
The current setting is displayed.
3. Select a filtering level.
– 0: no filtering
– 1: moderate filtering
– 2: high filtering
4. Confirm your selection.
You exit the menu automatically.
On non-calibrated scales, you can select the unit (Unit)
in which you want weight to be displayed.
(Unit)
CAUTION!
Patient hazard
In order to avoid misinterpretations, test results
for medical use must be displayed and used in
SI units only. Some devices offer the ability to
display test results in other units. This is only an
additional function.
– Use the results exclusively in SI units.
– The use of measurement results in non-SI
1. Select the item "Unit" from the menu.
2. Confirm your selection.
The current setting is displayed.
Select the unit in which you want weight to be
displayed:
– 9r: kilograms (kg)
– LbS: pounds (lbs)
– StS: stones (sts)
3. Confirm your selection.
You exit the menu automatically.
units is the sole responsibility of the user.

Publicité

loading